NPL is procuring a Josephson Arbitrary Waveform Synthesiser (JAWS) primary voltage standard. This is a series array of Josephson junctions in which single pulses of bias current each generate a small but precisely defined quantised voltage across the array. Accurate voltage waveforms are constructed by fast repetition of the pulses with accurately timed intervals between them. In the JAWS system, the output voltage is always precisely defined. This is in contrast to the Programmable Josephson Voltage Standard (PJVS) in which there are time intervals when the voltage output is not precisely known and is not suitable.
NPL intends to award the contract for a 2V JAWS reference system to the manufacturer the U.S. Department of Commerce, National Institute of Standards and Technology.
NPL requires the following specifications:
Range and accuracy of voltage output
Output voltage: Up to 2 V
Upper frequency limit: Up to at least 1 MHz
Lower frequency limit: Down to 5 Hz (lowest frequency component), with the capability to output DC
Pulse repetition rate: At least 10 Gsamples / second
RMS uncertainty with 100 Hz sine wave generation: better than 4 ppm (k=2)
RMS uncertainty with 10 kHz sine wave generation: better than 8 ppm (k=2)
Cooling
The system will not require any liquid cryogens
Automation
The system will be provided with software for custom waveform design or selecting standard sine waveforms. A detailed knowledge of the techniques used to generate the pulse patterns (e.g. delta-sigma conversion) should not be required of the operators. It is highly desirable that the software is able to perform automatic calibration of AC voltmeters and calibrators such as the Fluke 5700 series.
Training
The system should be simple enough to operate that users already experienced with programmable Josephson voltage standards can operate it after 2-3 days of training. The training will be provided by the supplier at the NPL Teddington site.
Preliminary market engagement notice (UK2). The buyer wants to talk to suppliers before designing the procurement. Taking part is how to shape it.
Bids are submitted through the buyer’s own portal, not on this site. Read the full notice on Find a Tender before relying on any date here.
No supplier information available.
This procurement was divided into 1 lots, each awarded separately.
| Lot | Value | Status |
|---|---|---|
| Lot 1 Feb 2026 – Jul 2026 | - | planning |