Supply and Installation of Wavemeter to the University of Birmingham
Wavelength meter capable of:
⢠Measurement of wavelengths between 413nm and 922nm (continuous, inclusive)
⢠Measurement accuracy of 2MHz
⢠Measurement resolution of <1MHz
⢠Calibration source either internal or provided by user
⢠User provided calibration source should be a free choice of wavelength within devices operating range
⢠Fibre coupled input using an FC/PC or FC/APC connector
⢠Device capable of being connected to a PC and providing data acquisition
⢠The device should be capable of simultaneously measuring 8 input channels or more
⢠The device should include the ability to generate error signal or PID feedback loops for 4 of the input channels or more, for stabilising these channel's wavelengths to a specific wavelength provided by the user.
⢠The device dimensions should be close to or less than 500mm x 400mm x 300mm
Overall budget for this item is up to £78,000 excluding VAT.
This project may be funded by the European Regional Development Fund (ERDF) or;
| Supplier | Identifier | Award Value | Cross-References |
|---|---|---|---|
| HighFinesse GmbH | - | £76,209 | - |