Requirement:
Cefas require to purchase a single Fluorescence Stereomicroscope with a high-resolution camera for microplastic work.
Technical requirement:
1. 1 x Fluorescence Stereomicroscope
2. Fluorescence LED illumination systems
3. Suitable for Nile Red
4. Automated XY sliding stage
5. High-Resolution camera and imaging software
The Fluorescence Stereomicroscope with High-Resolution Camera should be ideal for imaging Nile red stained microplastic fibers and fragments. The fluorescence filter should be GFP (excitation - 470/40x nm; emission 525/50nm). The microscope should include a sliding stage featuring XY movement control for the effective counting microplastic items and capturing of images. The camera system should be a high-resolution 20 MP camera and with imaging & video software.
| Supplier | Identifier | Award Value | Cross-References |
|---|---|---|---|
| GT Vision Ltd | 04974093 | £18,000 | - |
Government spending data: This supplier has received £171,732 in 14 payments (over £25k) from Manchester University NHS Foundation Trust, DEFRA, UKRI, Guy's and St Thomas' NHS FT, Humberside Police and 1 more public bodies (2023-02-03 to 2026-01-16). View full payment history →